HIMA Temperature Chamber
SOURCE
- HIMA Products, Dooravaningar, Bangalore
-
www.himaproduct.com
himaproduct@gmail.com -
+91-80-28513471/40968796
9845770531
-
QUICK REFERENCE
- Note Heating and Cooling cycles have to be done differently.
- The top part of the display is called
PVand the bottom part isSV. - Press the
SETbutton, thePVindicatesSTG(segment). eg. Choose0001 - Press
SETagain fordly(delay) and set the delay time in minutes.0000to start immediately. - Press
SETagain forset 1(First temperature setting). Each segment can have 10 settings. Use theupanddownkeys to set the Temperature Set Point 1. - Press
SETagain forrtme(Ramp time) in minutes - Press
SETagain for Hold time in minutes - Repeat this process for the next 9 set points, if you need them or you can just press enter to end the settings.
- And now you to press RESET to start the new settings else it will still be executing the old settings. And depending on the settings, the internal micro-controller may take a long time to calculate the control routine before starting. So have patience!